Anti-infective external coating of central venous catheters: a randomized, noninferiority trial comparing 5-fluorouracil with chlorhexidine/silver sulfadiazine in preventing catheter colonization.

Author(s): Walz JM, Avelar RL, Longtine KJ, Carter KL, Mermel LA, Heard SO

Affiliation(s): Department of Anesthesiology and Surgery, UMass Memorial Medical Center, Worcester, MA, USA. walzm@ummhc.org

Publication date & source: 2010-11, Crit Care Med., 38(11):2095-102.

Publication type: Comparative Study; Multicenter Study; Randomized Controlled Trial

OBJECTIVE: The antimetabolite drug, 5-fluorouracil, inhibits microbial growth. Coating of central venous catheters with 5-fluorouracil may reduce the risk of catheter infection. Our objective was to compare the safety and efficacy of central venous catheters externally coated with 5-fluorouracil with those coated with chlorhexidine and silver sulfadiazine. DESIGN: Prospective, single-blind, randomized, active-controlled, multicentered, noninferiority trial. SETTING: Twenty-five US medical center intensive care units. PATIENTS: A total of 960 adult patients requiring central venous catheterization for up to 28 days. INTERVENTIONS: Patients were randomized to receive a central venous catheter externally coated with either 5-fluorouracil (n = 480) or chlorhexidine and silver sulfadiazine (n = 480). MEASUREMENTS AND MAIN RESULTS: The primary antimicrobial outcome was a dichotomous measure (<15 colony-forming units or >/= 15 colony-forming units) for catheter colonization determined by the roll plate method. Secondary antimicrobial outcomes included local site infection and catheter-related bloodstream infection. Central venous catheters coated with 5-fluorouracil were noninferior to chlorhexidine and silver sulfadiazine coated central venous catheters with respect to the incidence of catheter colonization (2.9% vs. 5.3%, respectively). Local site infection occurred in 1.4% of the 5-fluorouracil group and 0.9% of the chlorhexidine and silver sulfadiazine group. No episode of catheter-related bloodstream infection occurred in the 5-fluorouracil group, whereas two episodes were noted in the chlorhexidine and silver sulfadiazine group. Only Gram-positive organisms were cultured from 5-fluorouracil catheters, whereas Gram-positive bacteria, Gram-negative bacteria, and Candida were cultured from the chlorhexidine and silver sulfadiazine central venous catheters. Adverse events were comparable between the two central venous catheter coatings. CONCLUSIONS: Our results suggest that central venous catheters externally coated with 5-fluorouracil are a safe and effective alternative to catheters externally coated with chlorhexidine and silver sulfadiazine when used in critically ill patients.

Page last updated: 2011-12-09

-- advertisement -- The American Red Cross
 
Home | About Us | Contact Us | Site usage policy | Privacy policy

All Rights reserved - Copyright DrugLib.com, 2006-2017